SCALE CALIBRATION (MODELS WITH A SCALE ONLY) 4 - 3
Step 1 - Electronic Entry into Calibration Mode
Encompasses all scales sold without a mechanical calibration switch installed. When the integrated scale
does not have a calibration switch, electronic entry is required.
Firmware Versions Older Than v3.4
To perform an electronic calibration entry:
Touch Scale Zero and Volume/Tone simultaneously for five consecutive seconds (see Features of the
MP6000 on page 1-9 for the location of these buttons on the front panel).
Firmware Versions Newer Than v3.4
To perform an electronic calibration entry:
Press and hold the Scale Zero and Volume buttons continuously for five seconds. After five seconds a
short beep sounds. Release the Scale Zero and Volume buttons.
Within two seconds after releasing the buttons, press the Scale Zero and Volume buttons again and
release. After release the system sounds five long beeps and enters scale calibration.
When Calibration Mode is entered successfully, the scan
ner emits five long beep sounds. The optional Scale
Display blinks CAL00 and CAL _ _ and the 7-segment display scrolls
C00Lb, or C00g. Once the scale
Calibration Mode is successfully entered, continue the Calibration process with Step 2 - Program Legal
Parameters on page 4-3.
Step 2 - Program Legal Parameters
Sealable and legal parameters (Unit Selection and Dampening Filter Setting) can only be
programmed/changed when the MP6000 is placed into Calibration Mode. There is no requirement to change
these values (they can remain in the default or present state). However, if there is a need to change their
values they must be programmed when the scale is in a Calibration Mode. The Dampening Filter Setting can
be changed multiple times during a Calibration Mode without affecting the calibration outcome.
After completing these changes, the calibration process re-starts with Step 3 - Calibration at NO LOAD on
page 4-4. If you change the Unit S
election value before successfully reaching Step 5 - Calibration Success or
Failure on page 4-7 and the Calibration Mode exits, the scale will be non-operational. The scale w
ill have to be
successfully calibrated before
it becomes operational.
Legal Scale Units (Unit Selection) - Kilograms or Pounds
To change the legal parameter Unit Selection (when the scale is in a Calibration Mode):
1. Verify the units of weight measurement by checking the Scale Display icon (the icon displays either lb or kg
units), or the 7-segment display which scrolls either
Lb or g units (g indicates kg units).
IMPORTANT If Calibration Mode was not entered successfully, the scale remains in its current state.