4 - 4 MP6000 INTEGRATOR GUIDE
2. If the weight unit needs to be changed (e.g., from kg units to lb units or vice-versa), scan the appropriate
Legal Scale Units bar code (see Legal Scale Units on page 4-13). If you do not change the default, or
current weight unit, the scale calibrates with the default (kg) or previously programmed unit.
Important Notes
1. Automatic zero setting is enabled for metric scales (kg), and disabled for US scales (lb).
2. Automatic zero setting, if enabled, is performed if weight has remained below zero, and stable (i.e., no
motion on the platter) for at least 5 seconds. The scale is automatically set to zero when the 5 seconds
expires.
3. If the weight units change from lb to kg, this setting is enabled after a cold power start of the MP6000. A
scale reset (e.g., via the Scale Reset STISCLRST bar code) WILL NOT enable this setting. Likewise, if
weight units change from kg to lb, this setting is disabled. This setting is illegal for US (lb) scales.
Legal Scale Dampening Filter
Set the scale sensitivity to vibration by scanning the appropriate bar code (see page 4-18). To program this
parameter, the scale must be in
Calibration Mode. Bar code command values are as follows:
•
0 = Higher (most sensitive to vibration)
•
1 = Low (factory default)
•
2 = Very Low
•
3 = Ultra Low
The lower the number, the more sensitive the scale is to vibr
ation. The higher the number, the slower the scale
may operate.
Step 3 - Calibration at NO LOAD
When Calibration Mode enters properly, the Scale Display blinks between CAL00 and CAL _ _ and the
7-segment display scrolls
C00Lb, or C00g. Scale calibration can begin.
1. Ensure the platter is installed and there is NO WEIGHTED LOAD on it.
2. Touch the Scale Zero button on the front panel.
3. If Calibration at NO WEIGHTED LOAD is successful after the Scale Zero button is pressed, the process
continues with Step 4 - Calibration at LOAD on page 4-5.
or
If Calibration at NO WEIGHTED LOAD is not successful after the Scale Zero button is pressed, the
process continues with Calibration Failure on page 4-7.
NOTE The unit selection legal parameter can be scanned at any time after entering a Calibration Mode.
After a successful scan, Calibration Mode restarts at Step 3 - Calibration at NO LOAD on page 4-4.
Additional scale parameter bar codes are located in the MP6000 Bar Code Programming Guide (p/n
72E-172633-xx).
IMPORTANT This step of the calibration procedure must be completed within 30 seconds, or 90 seconds for
Firmware Version before 3.4. (See Identifying Firmware Version.) If this time expires before
completing the step then the calibration procedure is exited.